Mohammad Mazloum-Ardakani; Azimeh Mandegari; Alireza Khoshroo Khoshroo; Saeed Masoum; Hadi Kargar
Volume 3, Issue 2 , September 2016, , Pages 96-104
Abstract
In this work, response surface methodology in conjunction with central composite design for modeling and optimization of the influence of some process variables (polyvinyl chloride (F1), ionophore (F2), additive (F3) and plasticizer (F4) amounts), on the performance of polyvinyl chloride membrane lead ...
Read More
In this work, response surface methodology in conjunction with central composite design for modeling and optimization of the influence of some process variables (polyvinyl chloride (F1), ionophore (F2), additive (F3) and plasticizer (F4) amounts), on the performance of polyvinyl chloride membrane lead (ІІ) ion-selective electrode is discussed. The slope of 29.1 ± 0.1 mV at the optimal amounts of polyvinyl chloride (0.0283 g), ionophore (0.0074 g), additive (0.002 g) and plasticizer (0.060 g) has been achieved. The electrode exhibited a linear potential response to lead (II) in the concentration range of 1.0 × 10-5 mol L-1 to 1.0 × 10-1 mol L-1 over pH range of 3.0 - 5.5. Greatly, the alternating current impedance technique was applied to investigate the response mechanism of the electrode. The results were obtained from electrochemical impedance spectroscopy shows a linear concentrations range of 1.0 × 10-6 mol L-1 to 1.0×10-1 mol L-1 and in comparison with potentiometry, the pH range increased to 2.5 − 6.0.